Diagnostic X-ray Tube for CT Scanners – Medical Radiology Equipment Accessory
SKU: UMY-XM-053
Brand: Umy
MOQ: 1
Origin: China
Quick Info.
- SKU NO.: UMY-XM-053
- Device Classification: Class Ⅱ
- Warranty: 1 Year
- Power Source: Electric
- Transport Package: Carton or Wooden Cases
- Origin: China
- Material: Aluminum and magnesium alloys
- After-Sale Service: Online Technical Support
- Production Capacity: 3000 Sets/Year
This diagnostic X-ray tube is designed as a high-quality accessory for CT scanners, enhancing imaging performance in medical radiology. Built for durability and precision, it provides consistent and accurate X-ray output crucial for detailed diagnostic scans. Ideal for hospitals and imaging centers, this X-ray tube ensures optimal functionality for CT systems, supporting reliable patient diagnoses. An essential component in radiology equipment, it meets the rigorous demands of modern diagnostic imaging.
The Specific Parameters
| ITEMS | PARAMETERS |
|---|---|
| Radiation Field Size (1 Meter from Focal Point) | 1129mm × 152mm |
| Tube Filtration | 1.8mm Al / 70kV |
| High Voltage Plug | Complies with GB/T 10151-2008 |
| Heat Capacity | 4.0 MHU (3.7MJ) |
| Cooling Method | Forced Fan Oil Cooling |
| Temperature Control Switch | Closed at 45~65ºC, Opened at 75~85ºC |
| Leakage Radiation in Loading State | ≤0.88mGy/h @ 140kV, 42mA |
| Nominal Anode Input Power (Small Focal Length Filament) | 28kW |
| Nominal Anode Input Power (Large Focal Length Filament) | 48kW |
| Maximum Continuous Heat Dissipation | 6kW |
| Weight | About 67kg |
| Product Model Specifications | XRT02 |
| Rotating Anode Rate | ≥6800 r/min @(120Hz) |
| Maximum Anode Heat Capacity | 4000KHU (2840kJ) |
| Anode Nominal Input Power | 28kW / 48kW |
| Focus Nominal Value f (YY/T0063) | 0.4 × 0.7, 0.6 × 1.3 |
| Exposure Rate | ≥0.03mC/kg·s |
| Anode Heating and Cooling Curve | Refer to Figure 4 |
| Nominal X-Ray Tube Voltage | 140kV |
| Target Disk Material | Rhenium Tungsten Molybdenum-Based Graphite Target |
| Target Angle | 7° |
| Inherent Filtration (IEC60522/1999) | <1.1mm Al / 70kV |
| Continuous Anode Input Power | 7kW |
| Filament Voltage | 3.0~12.5V / 6.0~16.5V |
| Filament Current | 3.0~5.5A |